RecallRadar

FDA Device recall Z-2364-2018

Zimmer¿ M/L Taper Hip Prosthesis With Kinectiv¿ Technology Cementless, Size 6

On 2018-07-11 the FDA classified a Class II recall of Zimmer¿ M/L Taper Hip Prosthesis With Kinectiv¿ Technology Cementless, Size 6 by Zimmer Biomet, citing M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be o….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2364-2018
ClassificationClass II
Statusterminated
Initiated2018-05-30
Published2018-07-11
Last updated2026-09-13 11:54 UTC
CompanyZimmer Biomet
Distributionn/a

Product

Zimmer¿ M/L Taper Hip Prosthesis With Kinectiv¿ Technology Cementless, Size 6; 00-7713-006-00

Reason

M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris.

Identifiers

code_info62044925 62571340 62831345 62869740 62953032 63068885 63081184 63089967

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.