FDA Device recall Z-2362-2026
Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXE…
- FDA Device
- Class II
- ongoing
- Published 2026-06-17
On 2026-06-17 the FDA classified a Class II recall of Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXE… by The Binding Site Group, citing it was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clin….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2362-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-04-16 |
| Published | 2026-06-17 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | The Binding Site Group |
| Distribution | US |
Product
Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.
Reason
It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.
Identifiers
| code_info | Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2362-2026%22