FDA Device recall Z-2361-2016

Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version…

On 2016-08-31 the FDA classified a Class I recall of Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version… by Alere San Diego, citing alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2361-2016
ClassificationClass I
Statusterminated
Initiated2016-07-11
Published2016-08-31
Last updated2026-09-13 11:54 UTC
CompanyAlere San Diego
DistributionUS

Product

Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor and disposable test strips. The monitor measures impedance, heats the test strip to the proper reaction temperature, and provides a user interface. The blood sample is applied to the Test Strip and the clotting reaction occurs on the Test Strip.

Reason

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Identifiers

code_infoAll Lots All Serial Numbers