FDA Device recall Z-2358-2016

Alere INRatio2 PT/INR Home Monitoring System Model Number: 0200432, 0200433 Product Usage: Alere INRatio 2 PT/INR Home Monitoring System: T…

On 2016-08-31 the FDA classified a Class I recall of Alere INRatio2 PT/INR Home Monitoring System Model Number: 0200432, 0200433 Product Usage: Alere INRatio 2 PT/INR Home Monitoring System: T… by Alere San Diego, citing alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2358-2016
ClassificationClass I
Statusterminated
Initiated2016-07-11
Published2016-08-31
Last updated2026-09-13 11:54 UTC
CompanyAlere San Diego
DistributionUS

Product

Alere INRatio2 PT/INR Home Monitoring System Model Number: 0200432, 0200433 Product Usage: Alere INRatio 2 PT/INR Home Monitoring System: The Alere INRatio 2 PT/INR Home Monitoring System, consisting of the INRatio 2 Home Monitor and INRatio PT/INR test strip, is used for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin therapy on clotting time by properly selected suitably trained users (by prescription for home use or other order of a treating physician). Patients must be stabilized (>6 weeks) on warfarin therapy. The Alere INRatio¿2 PT/INR Home Monitoring System is intended for use outside of the body (in vitro diagnostic use). The Alere INRatio¿2 PT/INR Home Monitoring System is not intended to be used for screening purposes.

Reason

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Identifiers

code_infoAll Lots All Serial Numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2358-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.