FDA Device recall Z-2354-2018

microTargeting" Lead Adapter for STar" Drive, Product Number 70-CN-DB

On 2018-07-11 the FDA classified a Class II recall of microTargeting" Lead Adapter for STar" Drive, Product Number 70-CN-DB by Fhc, citing the FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2354-2018
ClassificationClass II
Statusterminated
Initiated2018-05-10
Published2018-07-11
Last updated2026-09-13 11:54 UTC
CompanyFhc
DistributionAL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

microTargeting" Lead Adapter for STar" Drive, Product Number 70-CN-DB

Reason

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

Identifiers

code_infoUDI: 00873263002643

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2354-2018%22