On 2016-08-17 the FDA classified a Class II recall of VITROS Immunodiagnostic Products Estradiol Reagent Pack, Unique Device Identifier No. 10758750005017, REF/Catalog Number/Product Code 855 2… by Ortho Clinical Diagnostics, citing ortho Clinical Diagnostics has issued a recall of their VITROS immunodiagnostics Estradiol Reagent assay. The firm was made aware of a potential cross-reactivity between the drug….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2352-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-07-14 |
| Published | 2016-08-17 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ortho Clinical Diagnostics |
| Distribution | US |
Product
VITROS Immunodiagnostic Products Estradiol Reagent Pack, Unique Device Identifier No. 10758750005017, REF/Catalog Number/Product Code 855 2630; IVD -- Ortho-Clinical Diagnostics, Pencoed, Bridgend CF35 5PZ, UK --- The VITROS Estradiol test is performed using the VITROS Estradiol Reagent Pack and Calibrators on VITROS ECi/ECiQ Immunodiagnostic Systems, VITROS 3600 Immunodiagnostic Systems and VITROS 5600 Integrated Systems.
Reason
Ortho Clinical Diagnostics has issued a recall of their VITROS immunodiagnostics Estradiol Reagent assay. The firm was made aware of a potential cross-reactivity between the drug Fulvestrant and theVITROS Immunodiagnostic Estradiol Reagent assay. The firm conducted an internal investigation and was able to confirm biased positive (ranging from 1303-4046%) on Estradiol results, obtained from postmenopausal females taking Fulvestrant.
Identifiers
| code_info | Lot numbers: 1470 (expiry date 02-Aug-16); 1480 (expiry date 05-Aug-16), 1490 (expiry date 10-Oct-16), 1500 (expiry dat…Lot numbers: 1470 (expiry date 02-Aug-16); 1480 (expiry date 05-Aug-16), 1490 (expiry date 10-Oct-16), 1500 (expiry date 10-Oct-16), 1510 (expiry date 01-Nov-16), 1528 (expiry date 01-Nov-16) 1538 (expiry date 14-Mar-17), 1548 (expiry date 09-Mar-17), 1558 (expiry date 09-Mar-17) and 1568 (expiry date 29-Mar-17) |
|---|