FDA Device recall Z-2350-2019
ACUSON NX2 Diagnostic Ultrasound System, Model # 11284381, RX Only, Siemens Medical Solutions, containing software version VA10A, VA10B, VA…
- FDA Device
- Class II
- ongoing
- Published 2019-08-28
On 2019-08-28 the FDA classified a Class II recall of ACUSON NX2 Diagnostic Ultrasound System, Model # 11284381, RX Only, Siemens Medical Solutions, containing software version VA10A, VA10B, VA… by Siemens Medical Solutions Usa, citing the action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2350-2019 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2019-01-28 |
| Published | 2019-08-28 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
ACUSON NX2 Diagnostic Ultrasound System, Model # 11284381, RX Only, Siemens Medical Solutions, containing software version VA10A, VA10B, VA11A, VA11B; UDI # 04056869030005 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Reason
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware
Identifiers
| code_info | All product lots of the Acuson NX2, model # 11284381 containing the following software: VA10A, VA10B,VA10C, VA11A, VA11B, VA11C |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2350-2019%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.