FDA Device recall Z-2349-2019

Thoratec Heartmate 3 Sealed Outflow Graft with Bend Relief, Sterile, REF 105581US/105581, RX only, Part of Left Ventricular Assist Device (…

On 2019-08-28 the FDA classified a Class II recall of Thoratec Heartmate 3 Sealed Outflow Graft with Bend Relief, Sterile, REF 105581US/105581, RX only, Part of Left Ventricular Assist Device (… by Thoratec, citing outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2349-2019
ClassificationClass II
Statusterminated
Initiated2019-03-30
Published2019-08-28
Last updated2026-09-13 11:55 UTC
CompanyThoratec
DistributionUS

Product

Thoratec Heartmate 3 Sealed Outflow Graft with Bend Relief, Sterile, REF 105581US/105581, RX only, Part of Left Ventricular Assist Device (LVAD) UDI: 00813024013266

Reason

Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.

Identifiers

code_info
105581/105581US: The shelf life is 60 months from date of manufacture. All batch/lot numbers 6613000 and below or 5000…105581/105581US: The shelf life is 60 months from date of manufacture. All batch/lot numbers 6613000 and below or 50000000 and higher. UDI/GMDN 00813024013266 / 61747