RecallRadar

FDA Device recall Z-2349-2016

STAPLER 30, IS4000, model number 470430

On 2016-08-17 the FDA classified a Class II recall of STAPLER 30, IS4000, model number 470430 by Intuitive Surgical, citing an increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2349-2016
ClassificationClass II
Statusterminated
Initiated2016-07-14
Published2016-08-17
Last updated2026-09-13 11:54 UTC
CompanyIntuitive Surgical
DistributionPR

Product

STAPLER 30, IS4000, model number 470430; Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist Stapler 30 Instrument and Stapler 30 Reloads are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device is indicated for adult and pediatric use. The device can be used with staple line or tissue buttressing material.

Reason

An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System logs found an increased trend

Identifiers

code_infoAll

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2349-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.