RecallRadar

FDA Device recall Z-2348-2016

STAPLER 45, IS4000, model number 470298 Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist¿ Stapler 45, Stapler 45 Reloads…

On 2016-08-17 the FDA classified a Class II recall of STAPLER 45, IS4000, model number 470298 Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist¿ Stapler 45, Stapler 45 Reloads… by Intuitive Surgical, citing an increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2348-2016
ClassificationClass II
Statusterminated
Initiated2016-07-14
Published2016-08-17
Last updated2026-09-13 11:54 UTC
CompanyIntuitive Surgical
DistributionPR

Product

STAPLER 45, IS4000, model number 470298 Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist¿ Stapler 45, Stapler 45 Reloads and other Stapler Accessories are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device can be used with staple line or tissue butressing material (natural or synthetic).

Reason

An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System logs found an increased trend

Identifiers

code_infoAll

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.