RecallRadar

FDA Device recall Z-2348-2012

Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extrac…

On 2012-09-19 the FDA classified a Class II recall of Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extrac… by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2348-2012
ClassificationClass II
Statusterminated
Initiated2009-08-12
Published2012-09-19
Last updated2026-09-13 11:53 UTC
CompanyTerumo Cardiovascular Systems
Distributionn/a

Product

Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to warn when the blood level in the reservoir reaches the operator-set limit, and to stop the arterial and cardioplegia pumps.

Reason

Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav

Identifiers

code_infocatalog number 16404 and serial numbers: 4357, 4381, 4386, 4387, 4389, 4391, and 4392.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2348-2012%22