RecallRadar

FDA Device recall Z-2345-2021

Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device

On 2021-09-01 the FDA classified a Class II recall of Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device by Philips North America, citing pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2345-2021
ClassificationClass II
Statusterminated
Initiated2021-07-30
Published2021-09-01
Last updated2026-09-13 11:56 UTC
CompanyPhilips North America
DistributionUS

Product

Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 78127 (OUS)

Reason

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Identifiers

code_info
S/N and UDI: 78189 (01)00884838068452; 78272 (01)00884838068452; 78294 (01)00884838068452; 78295 (01)00884838068452; 78…S/N and UDI: 78189 (01)00884838068452; 78272 (01)00884838068452; 78294 (01)00884838068452; 78295 (01)00884838068452; 78332 (01)00884838068452; 78335 (01)00884838068452; 78340 (01)00884838068452; 78345 (01)00884838068452; 78249 (01)00884838068452; 78324 (01)00884838068452; 78326 (01)00884838068452