FDA Device recall Z-2344-2021

Ingenia Elition X-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device

On 2021-09-01 the FDA classified a Class II recall of Ingenia Elition X-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device by Philips North America, citing pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2344-2021
ClassificationClass II
Statusterminated
Initiated2021-07-30
Published2021-09-01
Last updated2026-09-13 11:56 UTC
CompanyPhilips North America
DistributionUS

Product

Ingenia Elition X-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781358

Reason

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Identifiers

code_info
S/N and UDI: US: 45245 (01)00884838088115(21)45245; 45247 (01)00884838088115(21)45247; 45261 (01)00884838088115(21)4526…S/N and UDI: US: 45245 (01)00884838088115(21)45245; 45247 (01)00884838088115(21)45247; 45261 (01)00884838088115(21)45261; 45263 (01)00884838088115(21)45263; 45264 (01)00884838088115(21)45264; 45269 (01)00884838088115(21)45269; OUS: 45215 (01)00884838088115; 45284 (01)00884838088115; 45253 (01)00884838088115; 45313 (01)00884838088115; 45265 (01)00884838088115; 45297 (01)00884838088115; 45218 (01)00884838088115; 45272 (01)00884838088115; 45281 (01)00884838088115; 45231 (01)00884838088115; 45278 (01)00884838088115; 45279 (01)00884838088115; 45245 (01)00884838088115; 45247 (01)00884838088115;