Olympus Endoscope Reprocessor OER-Mini-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and th…

On 2023-08-16 the FDA classified a Class II recall of Olympus Endoscope Reprocessor OER-Mini-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and th… by Olympus, citing revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2343-2023
ClassificationClass II
Statusongoing
Initiated2023-05-26
Published2023-08-16
Last updated2026-09-13 11:56 UTC
CompanyOlympus
DistributionUS

Product

Olympus Endoscope Reprocessor OER-Mini-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.

Reason

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

Identifiers

code_infoUDI-DI:- 04953170331619 Serial Numbers: 2300314 and the last three digits less than 314