FDA Device recall Z-2343-2016
91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B:…
- FDA Device
- Class II
- terminated
- Published 2016-08-10
On 2016-08-10 the FDA classified a Class II recall of 91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B:… by Spacelabs Healthcare, citing the firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is o….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2343-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-07-01 |
| Published | 2016-08-10 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Spacelabs Healthcare |
| Distribution | n/a |
Product
91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B: ECG/Respiration, Invasive Pressure x2, SpO2, Temperature x2 & adult/neonate NIBP. Option C: ECG/Respiration, Invasive Pressure x4, SpO2, Cardiac Output, Temperature x2 & adult/neonate NIBP. Option I: SpO2, Temperature x2, & adult/neonate NIBP.
Reason
The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data.
Identifiers
| code_info | Manifold Lot Codes: 128280, 129420, 129940, or 155640. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2343-2016%22