RecallRadar

FDA Device recall Z-2341-2015

Artis Q systems

On 2015-08-12 the FDA classified a Class II recall of Artis Q systems by Siemens Medical Solutions Usa, citing the possibility exists for a large amount of liquid to seep through the top cover of the table and penetrate the components inside the table. If this occurs, table movements may n….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2341-2015
ClassificationClass II
Statusterminated
Initiated2015-06-29
Published2015-08-12
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Artis Q systems; dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.

Reason

The possibility exists for a large amount of liquid to seep through the top cover of the table and penetrate the components inside the table. If this occurs, table movements may no longer be possible. No injuries reported.

Identifiers

code_info
Model Number of device: 10848282 10848281 10848355 10848354 serial numbers: 121006 121013 109016 121011 123008…Model Number of device: 10848282 10848281 10848355 10848354 serial numbers: 121006 121013 109016 121011 123008 123003 123011 121014 109011 121004 121018 121036 121012 109015 109009 121019 121010 121015 121016 109007 111000 121020 121009 121056 121057 109021 121008 109020 109027 109010 123010 121041

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2341-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.