FDA Device recall Z-2329-2020
STAINLESS STEEL REPROCESSING TRAY, 2D CAMERA HEAD, IS4000, Model No. 400499-03 - Product Usage: is designed to enable complex surgery using…
- FDA Device
- Class II
- terminated
- Published 2020-06-17
On 2020-06-17 the FDA classified a Class II recall of STAINLESS STEEL REPROCESSING TRAY, 2D CAMERA HEAD, IS4000, Model No. 400499-03 - Product Usage: is designed to enable complex surgery using… by Intuitive Surgical, citing small but detectable holes in the sterilization wrap used with the firm's 8 mm endoscope sterilization trays and handheld camera sterilization trays could lead to a potential brea….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2329-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-12-19 |
| Published | 2020-06-17 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Intuitive Surgical |
| Distribution | US |
Product
STAINLESS STEEL REPROCESSING TRAY, 2D CAMERA HEAD, IS4000, Model No. 400499-03 - Product Usage: is designed to enable complex surgery using a minimally invasive approach.
Reason
Small but detectable holes in the sterilization wrap used with the firm's 8 mm endoscope sterilization trays and handheld camera sterilization trays could lead to a potential breach of sterility and risk of infection. Certain features on the trays with handling and storage may cause these holes.
Identifiers
| code_info | Model No. 400499-03; All Batch Numbers; UDI: 00886874115855 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2329-2020%22