RecallRadar

FDA Device recall Z-2321-2012

AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com

On 2012-09-12 the FDA classified a Class II recall of AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com by Ameriwater, citing the firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water syste….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2321-2012
ClassificationClass II
Statusterminated
Initiated2012-08-09
Published2012-09-12
Last updated2026-09-13 11:53 UTC
CompanyAmeriwater
DistributionUS

Product

AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com. The AmeriWater MRO Portable Reverse Osmosis Systems are water treatment systems intended for use in hemodialysis applications. They are designed to pre-treat and purify potable water for use in making dialysate for hemodialysis and to meet current AAMI and Federal (U.S.) standards. The AmeriWater Portable MROS model is intended for use in a hospital, clinic, dialysis center, or for home care for single patient use. The AmeriWater Portable MRO1 model is for treatment of up to two patients in a hospital, clinic, or dialysis centers.

Reason

The firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water system. The incorrect fuse may result in excessive heat in the fuse holder eventually resulting in failure of the device to operate.

Identifiers

code_infoModel Number MROS, Catalog #00MROS20, 00MROS21 & Model Number MRO1, Catalog #00MRO120, and 00MRO121; Model Number(s): MROS and MRO1; Serial #'s 31000 to 31079.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2321-2012%22