728333 Spectral CT 7500 -Computed Tomography X-ray system

On 2021-08-25 the FDA classified a Class II recall of 728333 Spectral CT 7500 -Computed Tomography X-ray system by Philips North America, citing when setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased co….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2318-2021
ClassificationClass II
Statusterminated
Initiated2021-06-08
Published2021-08-25
Last updated2026-09-13 11:56 UTC
CompanyPhilips North America
DistributionUS

Product

728333 Spectral CT 7500 -Computed Tomography X-ray system

Reason

When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.

Identifiers

code_infoSerial Number/UDI: 10019 (01)00884838101111(21)10019 Software Version 5.0.0 SynchRight Option