FDA Device recall Z-2318-2017
Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001,…
- FDA Device
- Class II
- terminated
- Published 2017-06-14
On 2017-06-14 the FDA classified a Class II recall of Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001,… by Zimmer Biomet Spine, citing zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the clos….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2318-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-05-10 |
| Published | 2017-06-14 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Zimmer Biomet Spine |
| Distribution | US |
Product
Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017
Reason
Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.
Identifiers
| code_info | Model No: 07.02066.001 (All lots); 07.02053.001 (All lots distributed prior to April 26, 2017), and 730M0017 (Lot 309622) |
|---|