RecallRadar

FDA Device recall Z-2318-2017

Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle  90 in-lbs Model Number(s): 07.02066.001,…

On 2017-06-14 the FDA classified a Class II recall of Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle  90 in-lbs Model Number(s): 07.02066.001,… by Zimmer Biomet Spine, citing zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle  90 in-lbs because when applying torque to tighten the clos….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2318-2017
ClassificationClass II
Statusterminated
Initiated2017-05-10
Published2017-06-14
Last updated2026-09-13 11:54 UTC
CompanyZimmer Biomet Spine
DistributionUS

Product

Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle  90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017

Reason

Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle  90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.

Identifiers

code_infoModel No: 07.02066.001 (All lots); 07.02053.001 (All lots distributed prior to April 26, 2017), and 730M0017 (Lot 309622)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2318-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.