FDA Device recall Z-2315-2018
Diagnostic Ultrasound System, Aplio I-series
- FDA Device
- Class II
- terminated
- Published 2018-07-04
On 2018-07-04 the FDA classified a Class II recall of Diagnostic Ultrasound System, Aplio I-series by Toshiba American Medical Systems, citing the power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2315-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-01-04 |
| Published | 2018-07-04 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Toshiba American Medical Systems |
| Distribution | US |
Product
Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
Reason
The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
Identifiers
| code_info | Model No. TUS-AI700. TUS-AI800, TUS-AI900. Serial No. HNA1792011, HAB1732009, HNA1772007, HNA1772006, HAB1732023, HAB1…Model No. TUS-AI700. TUS-AI800, TUS-AI900. Serial No. HNA1792011, HAB1732009, HNA1772007, HNA1772006, HAB1732023, HAB1732013, HAB1732016, HAB16Z2002, NJC17X2086, NJC17X2087, JHA16X2004, HAB1732027, HAB16Z2001, HAB16Z2002, NJB1762003, JHA1752022, HAB16Z2003, HAB16Z2004, JHA1752019, NJB1782050, HAB1742033, NJB1772031, HNA1792013, HNA1772008, JHA1752010, HAB1742035, HAB1742036, NJB1762001, NJB1782042, JHA1762034, HAB1732030, HAB16Z2005, HAB1732007, HAB1742034, HAB1742037, JHA1752024, NJB1772019, JHA1762029, JHA1762031, NJB1782061, HAB1732019, HNA1772001, HAB1712008, HAB1722011, HAB1712010, HAB1712009, HNA1792014, HAB1722012, HAB16Z2003, HAB1732017, HAB1732018, HAB1712006, HAB1732014, HAB1712007, HAB1732015, HAB1732021, HAB1732020, HAB1732025, JHA1762030, HAB1732029, HAB1732031, NJC17X2066, HNA1772002, NJB1782060, NJB1782043, NJB1762002, JHA16X2002, NJB1782057, NJB1782055, JHA1752021, NJB1782058, NJB1762006, NJB1782056, NJC17X2089, NJB1772023, NJB1782051, NJB1782052, NJB1782053, NJB178205 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2315-2018%22