RecallRadar

FDA Device recall Z-2312-2012

Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended to be used by trained Nuclear Medicine Technologists to perform th…

On 2012-09-12 the FDA classified a Class II recall of Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended to be used by trained Nuclear Medicine Technologists to perform th… by Capintec, citing customers who received a Capintec Captus 3000 Thyroid Uptake System with Software Revisions 1.22, 1.23, 1.24, 1.25, and 1.26 may encounter an error while using the thyroid uptake….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2312-2012
ClassificationClass II
Statusterminated
Initiated2012-07-24
Published2012-09-12
Last updated2026-09-13 11:53 UTC
CompanyCapintec
DistributionUS

Product

Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended to be used by trained Nuclear Medicine Technologists to perform thyroid uptake procedures, which consist of the measurement of organ uptake and circulating radioactivity.

Reason

Customers who received a Capintec Captus 3000 Thyroid Uptake System with Software Revisions 1.22, 1.23, 1.24, 1.25, and 1.26 may encounter an error while using the thyroid uptake protocol ?Measure Liquid and Subtract Residual." This protocol was a new feature added to Software Revision 1.22 in January 2011. The error occurs only if the user does NOT measure the residual dose immediately after me

Identifiers

code_info
Catalog numbers for new systems: 5430-0076, 5430-0077, 5430-0103, and 5430-0087. Catalog numbers for conversion system…Catalog numbers for new systems: 5430-0076, 5430-0077, 5430-0103, and 5430-0087. Catalog numbers for conversion systems: 5430-0118 and 5430-0119. Serial numbers for new systems: 901006 through 901394, inclusive. Serial numbers for conversion systems: CNV 674 through CNV 730, inclusive. Also affected are devices which were serviced and devices for which customers received software upgrades between January 20, 2011 and June 30, 2012 with Revision 1.22, 1.23, 1.24, 1.25, or 1.26. (NOTE: Item number 5430-0075, Captus 3000 well counter system, is excluded. This system does not have the ability to perform thyroid uptake procedures.)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2312-2012%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.