RecallRadar

FDA Device recall Z-2309-2012

Zilver 518 Vascular Self Expanding Stent The Zilver Vascular Stent is intended for use as an adjunct to percutaneous transluminal angioplas…

On 2012-09-12 the FDA classified a Class II recall of Zilver 518 Vascular Self Expanding Stent The Zilver Vascular Stent is intended for use as an adjunct to percutaneous transluminal angioplas… by Cook Ireland, citing cook Ireland received a product complaint from Japan on December 22nd, 2011. The complainant reported that a 5 French Zilver Vascular Stent (part number ZIV5-18-125-8-80 from Lot#….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2309-2012
ClassificationClass II
Statusterminated
Initiated2012-01-05
Published2012-09-12
Last updated2026-09-13 11:53 UTC
CompanyCook Ireland
DistributionUS

Product

Zilver 518 Vascular Self Expanding Stent The Zilver Vascular Stent is intended for use as an adjunct to percutaneous transluminal angioplasty (PTA) in the treatment of symptomatic disease of iliac arteries up to 100 mm in length with a reference vessel diameter of 5 to 9 mm. Patients should be suitable candidates for PTA and / or stent treatment.

Reason

Cook Ireland received a product complaint from Japan on December 22nd, 2011. The complainant reported that a 5 French Zilver Vascular Stent (part number ZIV5-18-125-8-80 from Lot# CF694429) was inserted into a sheath over a wire guide where it became lodged in the sheath hub. After further examination, the user determined that the product was a 6 French Zilver Vascular Stent (part number ZIV6-35

Identifiers

code_infoPart order number: ZIV5-18-125-8-80; Lot Number: CF694429; Expiration Date: 09-2014

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2309-2012%22