RecallRadar

FDA Device recall Z-2307-2016

MultiDiagnost Eleva

On 2016-08-10 the FDA classified a Class II recall of MultiDiagnost Eleva by Philips Electronics North America, citing the two bolts that attach the clamping plate to the lateral shaft may break, which could cause the c-arm to move laterally forward, slide out of its bearings, and sag.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2307-2016
ClassificationClass II
Statusterminated
Initiated2016-07-19
Published2016-08-10
Last updated2026-09-13 11:54 UTC
CompanyPhilips Electronics North America
DistributionUS

Product

MultiDiagnost Eleva; x-ray system. Product Codes: 708032, 708036

Reason

The two bolts that attach the clamping plate to the lateral shaft may break, which could cause the c-arm to move laterally forward, slide out of its bearings, and sag

Identifiers

code_info
All serial numbers. 708032 System Serial numbers 100, 198, 354, 104, 367, 68, 428, 183, 414, 355, 289, 32, 269, 361,…All serial numbers. 708032 System Serial numbers 100, 198, 354, 104, 367, 68, 428, 183, 414, 355, 289, 32, 269, 361, 160, 423, 50, 56, 205, 350, 71, 208, 254, 108, 222, 286, 257, 253, 327, 128, 125, 40, 265, 174, 164, 214, 411, 263, 178, 103, 394, 425, 374, 284, 229, 79, 37, 11, 312, 116, 226, 296, 44, 49, 219, 207, 227, 154, 317, 18, 87, 406, 88, 244, 233, 315, 376, 288, 112, 402, 206, 261, 215, 31, 422, 344, 338, 342, 339, 384, 12, 136, 415, 416, 137, 228, 46, 126, 276, 280, 51, 193, 54, 232, 140, 20, 270, 121, 25, 59, 60, 48, 21, 28, 196, 199, 380, 316, 159, 153, 172, 14, 62, 216, 75, 282, 294, 340, 102, 76, 64, 13, 231, 106, 305, 372, 366, 364, 77, 189, 24, 279, 166, 221, 349, 72, 42, 424, 0, 80, 217, 43, 213, 197, 413, 145, 318, 210, 341, 111, 343, 55, 292, 241, 151, 246, 0, 110, 29, 141, 362, 35, 143, 236, 278, 99, 133, 365, 85, 267, 132, 134, 163, 127, 266, 412, 142, 144, 293, 358, 399, 303, 308, 238, 239, 287, 243, 74, 310, 34, 345, 39, 69, 409, 162, 240, 281, 368, 382, 357,

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2307-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.