FDA Device recall Z-2302-2020
Multi Display Manager in the Artis Systems (Image-Intensified Fluoroscopic X-Ray System) as follows: Artis Q floor Model: 10848280
- FDA Device
- Class II
- terminated
- Published 2020-06-17
On 2020-06-17 the FDA classified a Class II recall of Multi Display Manager in the Artis Systems (Image-Intensified Fluoroscopic X-Ray System) as follows: Artis Q floor Model: 10848280 by Siemens Medical Solutions Usa, citing artis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2302-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-04-30 |
| Published | 2020-06-17 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
Multi Display Manager in the Artis Systems (Image-Intensified Fluoroscopic X-Ray System) as follows: Artis Q floor Model: 10848280; Artis Q ceiling Model: 10848281; Artis Q biplane Model:10848282; Artis zee ceiling Model: 10094137; Artis zee floor Model: 10094135 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Reason
Artis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical contact, and result in planned procedures to be terminated and performed on an alternative x-ray system
Identifiers
| code_info | Serial Numbers: Serial 125253 103340 121526 121521 103335 109621 138311 148418 147237 109620 136555 121520 121525 121518 125254 |
|---|
Status history