FDA Device recall Z-2302-2016
VITROS Immunodiagnostic Products Total T4 Reagent Pack, REF/Product Code 874 4468, Unique Identifier No. 10758750005147, IVD --- Ortho-Clin…
- FDA Device
- Class II
- terminated
- Published 2016-08-10
On 2016-08-10 the FDA classified a Class II recall of VITROS Immunodiagnostic Products Total T4 Reagent Pack, REF/Product Code 874 4468, Unique Identifier No. 10758750005147, IVD --- Ortho-Clin… by Ortho Clinical Diagnostics, citing ortho-Clinical Diagnostics confirmed that sample results generated from VITROS Immunodiagnostic Products Total T4 Reagent Packs may show a negative bias prior to the open pack sto….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2302-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-06-02 |
| Published | 2016-08-10 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ortho Clinical Diagnostics |
| Distribution | US |
Product
VITROS Immunodiagnostic Products Total T4 Reagent Pack, REF/Product Code 874 4468, Unique Identifier No. 10758750005147, IVD --- Ortho-Clinical Diagnostics, For in vitro diagnostic use only. For the quantitative measurement of total thyroxine (T4) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System to aid in the differential diagnosis of thyroid disease.
Reason
Ortho-Clinical Diagnostics confirmed that sample results generated from VITROS Immunodiagnostic Products Total T4 Reagent Packs may show a negative bias prior to the open pack storage limit of <= 8 weeks that is located in the Instructions for Use (IFU) for VITROS Immunodiagnostic Products Total T4 Reagent Pack.
Identifiers
| code_info | Lot Numbers:1836, 1840, 1845, 1850, 1860, 1870, 1880, 1890, 1900, 1910, 1920, 1930, 1940, 1950, 1968, 1978, 1988, 1998 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2302-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.