FDA Device recall Z-2299-2020

Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ Aspiration System, UDI: 00850291007079

On 2020-06-17 the FDA classified a Class II recall of Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ Aspiration System, UDI: 00850291007079 by Inari Medical, citing the hemostasis valve of aspiration guide catheter devices may fail to seal resulting in a loss of vacuum during thrombus aspiration, which may result the procedure being prolonged….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2299-2020
ClassificationClass II
Statusterminated
Initiated2020-03-23
Published2020-06-17
Last updated2026-09-13 11:55 UTC
CompanyInari Medical
DistributionAL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MO, MS, NC, ND, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI

Product

Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ Aspiration System, UDI: 00850291007079

Reason

The hemostasis valve of aspiration guide catheter devices may fail to seal resulting in a loss of vacuum during thrombus aspiration, which may result the procedure being prolonged as another device must be used or suboptimal thrombus removal if the problem is not detected.

Identifiers

code_infoLot: 20010004, 20010005, 20010008

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.