FDA Device recall Z-2298-2016
Syva¿ EMIT¿ 2000 Theophylline - Catalog # 4P019UL /SMN#10445324 is a homogeneous enzyme immunoassay intended for use in the quantitative an…
- FDA Device
- Class II
- terminated
- Published 2016-08-03
On 2016-08-03 the FDA classified a Class II recall of Syva¿ EMIT¿ 2000 Theophylline - Catalog # 4P019UL /SMN#10445324 is a homogeneous enzyme immunoassay intended for use in the quantitative an… by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics has confirmed the Syva¿ EMIT¿ 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA¿ Chemistry Theophy….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2298-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-06-17 |
| Published | 2016-08-03 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | CA, CT, DE, FL, GA, IA, IN, KS, KY, LA, MA, MD, MI, MO, NC, NJ, NM, NY, OH, OK, PA, TN, VA, WA |
Product
Syva¿ EMIT¿ 2000 Theophylline - Catalog # 4P019UL /SMN#10445324 is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma.
Reason
Siemens Healthcare Diagnostics has confirmed the Syva¿ EMIT¿ 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA¿ Chemistry Theophylline_2 (THEO_2) reagent lot 334018, may exhibit increased imprecision for commercially available Bio-Rad Liquichek TDM Quality Control Levels and patient samples.
Identifiers
| code_info | Syva¿ EMIT¿ 2000 Theophylline - Catalog # 4P019UL /SMN#10445324 Syva¿ EMIT¿ 2000 Theophylline - kit lot G3 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2298-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.