FDA Device recall Z-2295-2016
Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non steri…
- FDA Device
- Class II
- terminated
- Published 2016-08-03
On 2016-08-03 the FDA classified a Class II recall of Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non steri… by Draeger Medical, citing risk associated with operating the device in skin temperature mode with temperature sensors attached but no patient connected.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2295-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-06-22 |
| Published | 2016-08-03 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Draeger Medical |
| Distribution | US |
Product
Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non sterile, software controlled. Product Usage: Babytherm 8004/8010 is an open care infant warmer system for warming premature babies, neonates and infants with a body weight of up to 8 kg (17.6 lbs). The device is intended for use in delivery rooms, operating rooms, maternity and obstetric units, neonatal and pediatric wards, neonatal and pediatric intensive care units. Therapy and nursing uses: - As a resuscitation table and in birthing rooms, obstetric units, and NICUs - As infant warmer bassinet during normal and intensive care - For infant re-warming and compensation of heat loss - For thermal stabilizing of infants with mattress heater (Babytherm 8010 only) - For gentle warming therapy using the combination of radiant and mattress heater - Cooling infants with fever (temperature of mattress below infant body temperature, Babytherm 8010 only) - With adjustable bed tilt for treatment of infants in Trendelenburg and reverse Trendelenburg position - For weaning infants from incubators - For O2 and nebulizer therapy (with appropriate accessories) - For lowering an infant s bilirubin levels when using the phototherapy option
Reason
Risk associated with operating the device in skin temperature mode with temperature sensors attached but no patient connected.
Identifiers
| code_info | Serial Numbers: ARLN-xxxx to ASFE-xxxx Catalog Numbers: FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2295-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.