Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical laboratories

On 2019-08-28 the FDA classified a Class II recall of Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical laboratories by Luminex, citing an incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients speci….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2292-2019
ClassificationClass II
Statusterminated
Initiated2018-06-29
Published2019-08-28
Last updated2026-09-13 11:55 UTC
CompanyLuminex
DistributionUS

Product

Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical laboratories. Scanner Label/Labeling: JADAK JDK-2330 REV B 170512-017

Reason

An incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient order.

Identifiers

code_info
ARIES System UDI: 00840487101537 ARIES M1 System UDI: 00840487100080 ARIES-M12V1-IVD, ARIES System ARIES-M6V1-…ARIES System UDI: 00840487101537 ARIES M1 System UDI: 00840487100080 ARIES-M12V1-IVD, ARIES System ARIES-M6V1-IVD, ARIES M1 System CN-0321-01, ARIES System Handheld Barcode Scanner All scanners manufactured after April 2017 and including Part Number JDK-2330 Rev B.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2292-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.