FDA Device recall Z-2291-2026

Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro So…

On 2026-06-10 the FDA classified a Class II recall of Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro So… by Philips Medical Systems Nederland B V, citing during the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly to….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2291-2026
ClassificationClass II
Statusongoing
Initiated2023-12-07
Published2026-06-10
Last updated2026-09-13 11:58 UTC
CompanyPhilips Medical Systems Nederland B V
DistributionUS

Product

Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Reason

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

Identifiers

code_info
UDI (01)00884838099258(21)450, (01)00884838099258(21)750, (01)00884838099258(21)409, (01)00884838099258(21)262, (01)008…UDI (01)00884838099258(21)450, (01)00884838099258(21)750, (01)00884838099258(21)409, (01)00884838099258(21)262, (01)00884838099258(21)275, (01)00884838099258(21)277, (01)00884838099258(21)295, (01)00884838099258(21)272, (01)00884838099258(21)261, (01)00884838099258(21)630, (01)00884838099258(21)1003, (01)00884838099258(21)1315, (01)00884838099258(21)381, (01)00884838099258(21)959, (01)00884838085268(21)1013, (01)00884838085268(21)1412, (01)00884838099258(21)232, (01)00884838085268(21)460, (01)00884838085268(21)1387, (01)00884838099258(21)260, (01)00884838099258(21)1019, (01)00884838099258(21)1371, (01)00884838085268(21)40, (01)00884838085268(21)43, (01)00884838085268(21)35, (01)00884838099258(21)1049, (01)00884838099258(21)1484, (01)00884838099258(21)1083, (01)00884838085268(21)703307, (01)00884838099258(21)246, (01)00884838099258(21)475.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2291-2026%22