FDA Device recall Z-2288-2018
HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160
- FDA Device
- Class II
- terminated
- Published 2018-07-04
On 2018-07-04 the FDA classified a Class II recall of HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160 by Cenorin, citing there is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2288-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-04-09 |
| Published | 2018-07-04 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Cenorin |
| Distribution | US |
Product
HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160; Model 610HT, 115V & Model 610HT, 230V Product Usage: Clean and thermally disinfect medical devices using full immersion pasteurization
Reason
There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.
Identifiers
| code_info | Serial Numbers: 611503, 611504, 611505, 611506, 611601, 611602, 611603, 611604, 611605, 611606, 611607, 611608, 611609 & 611610 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2288-2018%22