FDA Device recall Z-2286-2026
Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software k…
- FDA Device
- Class II
- ongoing
- Published 2026-06-10
On 2026-06-10 the FDA classified a Class II recall of Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software k… by Philips Medical Systems Nederland B V, citing during the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly to….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2286-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-12-07 |
| Published | 2026-06-10 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Philips Medical Systems Nederland B V |
| Distribution | US |
Product
Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Reason
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Identifiers
| code_info | UDI (01)00884838099227(21)88, (01)00884838099227(21)135, (01)00884838099227(21)63. |
|---|