RecallRadar

FDA Device recall Z-2286-2026

Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software k…

On 2026-06-10 the FDA classified a Class II recall of Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software k… by Philips Medical Systems Nederland B V, citing during the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly to….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2286-2026
ClassificationClass II
Statusongoing
Initiated2023-12-07
Published2026-06-10
Last updated2026-09-13 11:58 UTC
CompanyPhilips Medical Systems Nederland B V
DistributionUS

Product

Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Reason

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

Identifiers

code_infoUDI (01)00884838099227(21)88, (01)00884838099227(21)135, (01)00884838099227(21)63.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2286-2026%22