RecallRadar

FDA Device recall Z-2284-2012

QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor

On 2012-09-05 the FDA classified a Class II recall of QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor by Steris, citing STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2284-2012
ClassificationClass II
Statusterminated
Initiated2012-08-10
Published2012-09-05
Last updated2026-09-13 11:53 UTC
CompanySteris
DistributionUS

Product

QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.

Reason

STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..

Identifiers

code_infoLot numbers: 1879717, 1943786, 2054914, 2143717, 2177723, 2342319, 2730752, 2922565, 3094190

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2284-2012%22