FDA Device recall Z-2282-2020
ADVIA Chemistry Enzymatic Creatinine_2, Siemens Material Number 10335869, UDI Number 00630414526690 - Product Usage: use in the quantitativ…
- FDA Device
- Class II
- terminated
- Published 2020-06-17
On 2020-06-17 the FDA classified a Class II recall of ADVIA Chemistry Enzymatic Creatinine_2, Siemens Material Number 10335869, UDI Number 00630414526690 - Product Usage: use in the quantitativ… by Siemens Healthcare Diagnostics, citing the presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2282-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-05-04 |
| Published | 2020-06-17 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
ADVIA Chemistry Enzymatic Creatinine_2, Siemens Material Number 10335869, UDI Number 00630414526690 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.
Reason
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Identifiers
| code_info | Lot #s 53149, 53150, 53151, 53152, 53941, 53942, 53943, 53944, 54354, 54371, 54372, 54373, 54374 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2282-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.