RecallRadar

FDA Device recall Z-2280-2020

ADVIA Chemistry Fructosamine Reagent (FRUC), Siemens Material Number 10361941, UDI Number 00630414558073 - Product Usage: use in the quanti…

On 2020-06-17 the FDA classified a Class II recall of ADVIA Chemistry Fructosamine Reagent (FRUC), Siemens Material Number 10361941, UDI Number 00630414558073 - Product Usage: use in the quanti… by Siemens Healthcare Diagnostics, citing the presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2280-2020
ClassificationClass II
Statusterminated
Initiated2020-05-04
Published2020-06-17
Last updated2026-09-13 11:55 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

ADVIA Chemistry Fructosamine Reagent (FRUC), Siemens Material Number 10361941, UDI Number 00630414558073 - Product Usage: use in the quantitative measurement of glycated protein (fructosamine) in human serum and plasma for monitoring the blood glucose levels of diabetic patients over the preceding two-three weeks.

Reason

The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.

Identifiers

code_infoLot # 501874

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2280-2020%22