RecallRadar

FDA Device recall Z-2279-2019

BD Vacutainer Fluoride Tubes for Blood Alcohol determinations Part Number: 367001 Product Usage: Blood Alcohol determinations

On 2019-08-21 the FDA classified a Class II recall of BD Vacutainer Fluoride Tubes for Blood Alcohol determinations Part Number: 367001 Product Usage: Blood Alcohol determinations by Becton Dickinson And, citing lot has been confirmed to have no additive within the tube. Sample processed without the preservative (additive) in the tube, testing has yielded reliable results if the samples w….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2279-2019
ClassificationClass II
Statusterminated
Initiated2019-05-30
Published2019-08-21
Last updated2026-09-13 11:55 UTC
CompanyBecton Dickinson And
DistributionUS

Product

BD Vacutainer Fluoride Tubes for Blood Alcohol determinations Part Number: 367001 Product Usage: Blood Alcohol determinations

Reason

Lot has been confirmed to have no additive within the tube. Sample processed without the preservative (additive) in the tube, testing has yielded reliable results if the samples were stored at room temperature for no longer than two days. If the sample was stored for more than 2 days, the result for blood alcohol determination might not be accurate (either falsely low or falsely high)

Identifiers

code_infoLot Number: 8187663 Exp. Date: 2020/7/31 UDI: (01)30382903670018 (17)200731(10)8187663 (30)0100

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2279-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.