FDA Device recall Z-2277-2025
Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135
- FDA Device
- Class II
- ongoing
- Published 2025-08-13
On 2025-08-13 the FDA classified a Class II recall of Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135 by Stryker, citing tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. O….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2277-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-07-02 |
| Published | 2025-08-13 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Stryker |
| Distribution | US |
Product
Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP 24X4,1BLA, 2PRT QUICK / 5921-024-235
Reason
Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.
Identifiers
| code_info | 5921-018-135 GTIN 07613154599257 5921-018-235 GTIN 07613154599295 5921-024-235 GTIN 07613154599370 |
|---|