FDA Device recall Z-2273-2023
ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery
- FDA Device
- Class II
- ongoing
- Published 2023-08-09
On 2023-08-09 the FDA classified a Class II recall of ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery by Maquet Cardiovascular, citing the Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and posit….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2273-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-06-28 |
| Published | 2023-08-09 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Maquet Cardiovascular |
| Distribution | US |
Product
ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery.
Reason
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
Identifiers
| code_info | UDI-DI: 00607567700536 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2273-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.