FDA Device recall Z-2265-2012
Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are…
- FDA Device
- Class II
- terminated
- Published 2012-09-05
On 2012-09-05 the FDA classified a Class II recall of Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are… by Canon, citing screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2265-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2010-11-01 |
| Published | 2012-09-05 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Canon |
| Distribution | US |
Product
Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.
Reason
Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.
Identifiers
| code_info | Catalog numbers 3930B002AA, 3928B002AA, and 4066B002AA. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2265-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.