FDA Device recall Z-2264-2024

DxI 9000 Access Immunoassay Analyzer, C11137

On 2024-07-10 the FDA classified a Class II recall of DxI 9000 Access Immunoassay Analyzer, C11137 by Beckman Coulter, citing beckman Coulter is recalling their DxI 9000 Access Immunoassay Analyzers, an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2264-2024
ClassificationClass II
Statusongoing
Initiated2024-04-18
Published2024-07-10
Last updated2026-09-13 11:57 UTC
CompanyBeckman Coulter
Distributionn/a

Product

DxI 9000 Access Immunoassay Analyzer, C11137; in-vitro diagnostic

Reason

Beckman Coulter is recalling their DxI 9000 Access Immunoassay Analyzers, an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids, by correction. The reason for the recall is: Beckman Coulter has determined that the DxI 9000 Access Immunoassay analyzer accepts calibrator values scanned from multiple calibrator cards when an operator configures a new calibrator lot. The analyzer accepts the calibrator lot values from any scanned calibrator card, regardless of the assay test code or calibrator material lot number. The software is not validating the integrity of each calibrator barcode scanned to ensure they are all from the same calibrator card. This includes all software versions and all DxI 9000 instruments that are currently in the field. This issue does not impact calibrator material with 2D calibrator card barcodes. This may cause a failed calibration curve. The health risk to patients is dependent upon the analytes that could be impacted by the error, to include receiving delayed or erroneous results. The issue was identified by Beckman Coulter while conducting a field action on DxI 9000 Access Immunoassay Analyzers.

Identifiers

code_info
UDI/DI 15099590230623, Serial Numbers: 300116, 300117, 300123, 300124, 300126, 300128, 300129, 300130, 300131, 300133,…UDI/DI 15099590230623, Serial Numbers: 300116, 300117, 300123, 300124, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300137, 300138, 300139, 300141, 300142, 300143, 300145, 300146, 300147, 300148, 300149, 300150, 300151, 300152, 300153, 300154, 300155, 300156, 300157, 300158, 300160, 300161, 300162, 300163, 300164, 300166, 300167, 300168, 300169, 300170, 300171, 300172, 300174, 300176, 300178, 300179, 300180, 300181, 300182, 300183, 300188, 300189, 300190, 300191, 300192, 300193, 300194, 300195, 300196, 300197, 300198, 300199, 300200, 300201, 300202, 300203, 300204, 300205, 300206, 300207, 300208, 300209, 300210, 300211, 300212, 300213, 300214, 300215, 300216, 300217, 300218, 300219, 300220, 300221, 300222, 300223, 300224, 300226, 300227, 300228, 300229, 300230, 300231, 300232, 300233, 300234, 300235, 300236, 300237, 300238, 300239, 300240, 300241, 300242, 300243, 300244, 300245, 300246, 300247, 300248, 300249, 300251, 300252, 300253, 300254, 300255, 300256,

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2264-2024%22