RecallRadar

FDA Device recall Z-2264-2012

syngo Imaging XS. Radiological image processing system

On 2012-09-05 the FDA classified a Class II recall of syngo Imaging XS. Radiological image processing system by Siemens Medical Solutions Usa, citing there is a potential malfunction when using syngo Imaging XS, version VA70A or higher. In some situations it can happen that a miscalculation is done for the grey scale values in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2264-2012
ClassificationClass II
Statusterminated
Initiated2012-07-18
Published2012-09-05
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

syngo Imaging XS. Radiological image processing system.

Reason

There is a potential malfunction when using syngo Imaging XS, version VA70A or higher. In some situations it can happen that a miscalculation is done for the grey scale values in the functions "Regions of Interest", "Pixel Lens", "Edge Enhancement" and "Histograms".

Identifiers

code_infoModel number 10496279

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2264-2012%22