Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, T…
- FDA Device
- Class I
- terminated
- Published 2020-06-17
On 2020-06-17 the FDA classified a Class I recall of Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, T… by Endologix, citing A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2263-2020 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2020-05-06 |
| Published | 2020-06-17 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Endologix |
| Distribution | US |
Product
Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
Reason
A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.
Identifiers
| code_info | All Lot and Serial Numbers not yet implanted |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.