FDA Device recall Z-2263-2013

Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris.com

On 2013-10-02 the FDA classified a Class II recall of Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris.com by Steris, citing the air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the m….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2263-2013
ClassificationClass II
Statusterminated
Initiated2013-08-29
Published2013-10-02
Last updated2026-09-13 11:53 UTC
CompanySteris
DistributionUS

Product

Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris.com. The intended use of the REPS is washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-resistant, semi-critical devices.

Reason

The air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the main air/water channel within the light guide (LG) connection head and, as a result, the air pipe does not receive forced flow during the Reliance EPS processing cycle.

Identifiers

code_infoModel #'s: MB30001, MB30002, MB30003, MB000077, MB000078, MB000079, MB000080 and MB000081. Serial number range 3625606001 - 3608912002

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.