Always-On Tip Tracked Instruments (SPiN Drive¿ instruments) Brush, 15mm L, 1.8mm OD Intended for use in clinical interventions and for anat…

On 2023-08-09 the FDA classified a Class II recall of Always-On Tip Tracked Instruments (SPiN Drive¿ instruments) Brush, 15mm L, 1.8mm OD Intended for use in clinical interventions and for anat… by Olympus, citing electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2262-2023
ClassificationClass II
Statusongoing
Initiated2023-05-15
Published2023-08-09
Last updated2026-09-13 11:56 UTC
CompanyOlympus
DistributionUS

Product

Always-On Tip Tracked Instruments (SPiN Drive¿ instruments) Brush, 15mm L, 1.8mm OD Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic bronchoscopy are currently used for visualizing Model: INS-0352

Reason

Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay

Identifiers

code_infoUDI-DI: 00815686020538 Lot Numbers: 04865230201 and below

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2262-2023%22