RecallRadar

FDA Device recall Z-2262-2018

Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17

On 2018-06-27 the FDA classified a Class II recall of Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17 by Trimed, citing during a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2262-2018
ClassificationClass II
Statusterminated
Initiated2018-04-13
Published2018-06-27
Last updated2026-09-13 11:54 UTC
CompanyTrimed
DistributionCA, FL, IL, KS, MN, MO, MS, NC, NJ, NV, PA, TX, UT, WA

Product

Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19

Reason

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Identifiers

code_info
Lot #: a) R901, 03UU, 050522, 10505-22, 0344, 0051, 11805, 00AJ, 10505-23, 050523, 180502, 180901, 180503…Lot #: a) R901, 03UU, 050522, 10505-22, 0344, 0051, 11805, 00AJ, 10505-23, 050523, 180502, 180901, 180503, 18902; b) 03UV, 050523, 10505-23, 00AJ, 0052, 180901, 11809, 180902, 26408, 264-08, 11805, 180502, 118901

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2262-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.