Dual Chamber Container fill and transfer sets, 250 mL / 1250 mL or 500 mL / 2500 mL

On 2016-08-03 the FDA classified a Class II recall of Dual Chamber Container fill and transfer sets, 250 mL / 1250 mL or 500 mL / 2500 mL by The Metrix, citing an issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2262-2016
ClassificationClass II
Statusterminated
Initiated2016-06-23
Published2016-08-03
Last updated2026-09-13 11:54 UTC
CompanyThe Metrix
DistributionAL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, WA

Product

Dual Chamber Container fill and transfer sets, 250 mL / 1250 mL or 500 mL / 2500 mL

Reason

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.

Identifiers

code_info
All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique f…All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence. Affected product codes for this product: 58733 and 58734