FDA Device recall Z-2261-2020

Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of adult, pediatric and neon…

On 2020-06-10 the FDA classified a Class II recall of Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of adult, pediatric and neon… by Spacelabs Healthcare, citing the firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm co….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2261-2020
ClassificationClass II
Statusterminated
Initiated2019-08-30
Published2020-06-10
Last updated2026-09-13 11:55 UTC
CompanySpacelabs Healthcare
DistributionUS

Product

Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.

Reason

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

Identifiers

code_infoModel No. 96102 Xhibit Central; Software Versions: 1.3.1, 1.3.2, 1.3.3 and 1.3.4; UDI: 10841522100345

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.