RecallRadar

FDA Device recall Z-2255-2024

BD Pyxis, Name/REF: MEDBANK MINI 1FH-1FM/ 169-137, MEDBANK MINI 1FH-2HH MEDPASS/ 169-151, MEDBANK MINI 1HH-1FH-1HM/ 169-101, MEDBANK MINI 1…

On 2024-07-10 the FDA classified a Class II recall of BD Pyxis, Name/REF: MEDBANK MINI 1FH-1FM/ 169-137, MEDBANK MINI 1FH-2HH MEDPASS/ 169-151, MEDBANK MINI 1HH-1FH-1HM/ 169-101, MEDBANK MINI 1… by Carefusion 303, citing multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical st….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2255-2024
ClassificationClass II
Statusongoing
Initiated2024-05-13
Published2024-07-10
Last updated2026-09-13 11:57 UTC
CompanyCarefusion 303
DistributionUS

Product

BD Pyxis, Name/REF: MEDBANK MINI 1FH-1FM/ 169-137, MEDBANK MINI 1FH-2HH MEDPASS/ 169-151, MEDBANK MINI 1HH-1FH-1HM/ 169-101, MEDBANK MINI 1HH-1HM-1FM/ 169-102, MEDBANK MINI 1HH-3HM/ 169-103, MEDBANK MINI 2FH/ 169-104, MEDBANK MINI 2FM/ 169-105, MEDBANK MINI 2HH-1FH/ 169-106, MEDBANK MINI 2HH-1FM/ 169-107, MEDBANK MINI 2HH-1FM MEDPASS/ 169-152, MEDBANK MINI 2HH-2HM/ 169-108, MEDBANK MINI 2HM-1FM/ 169-109, MEDBANK MINI 3HH-1FG/ 169-110, MEDBANK MINI 3HH-1HM/ 169-111, MEDBANK MINI 4HH/ 169-112, MEDBANK MINI 4HH MEDPASS/ 169-150, MEDBANK MINI 4HM/ 169-113, MEDBANK MINI CR-2HH-1FH-P/ 169-114, MEDBANK MINI CR-4HH-P/ 169-115, MB MINI MAIN CR/ 169-163, MB MINI MAIN W/CUBIE DRWS/ 169-159, MB MINI MAIN W/FH FALSE DRW/ 169-162, MB MINI MAIN W/HH FALSE DRW/ 169-161, MB MINI MAIN W/MATRIX DRWS / 169-160, MB MINI MAIN W/MP/ 169-164 Used with the following user guides: BD Pyxis CUBIE Replenishment Station User Guide (DIR 10000414989), BD Pyxis MedBank Cabinet User Guide (DIR 10000414987), BD Pyxis MedBank Installation and Upgrade Guide (DIR 10000414990)

Reason

Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.

Identifiers

code_infoUDI-DI: 10885403512544. Devices Manufactured November 15, 2012 - July, 25 2023 User Guide, DIR 10000414989/DIR 10000414987/DIR 10000414990, Version 00.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2255-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.