BD Trucount Tubes: Catalog number: 340334 Hematology: BD Trucount tubes are used for determining absolute counts of leucocytes in blood

On 2016-07-27 the FDA classified a Class II recall of BD Trucount Tubes: Catalog number: 340334 Hematology: BD Trucount tubes are used for determining absolute counts of leucocytes in blood by Becton Dickinson And Bd Biosciences, citing one lot of BD Trucount Absolute Counting Tubes (IVD) may not be sealed properly and exposure to ambient environment can rapidly degrade the performance of the product.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2252-2016
ClassificationClass II
Statusterminated
Initiated2016-06-24
Published2016-07-27
Last updated2026-09-13 11:54 UTC
CompanyBecton Dickinson And Bd Biosciences
DistributionAR, AZ, CA, IN, KY, MA, MD, NC, NJ, NY, OH, PA, PR, SD, VA, WA, WI

Product

BD Trucount Tubes: Catalog number: 340334 Hematology: BD Trucount tubes are used for determining absolute counts of leucocytes in blood.

Reason

One lot of BD Trucount Absolute Counting Tubes (IVD) may not be sealed properly and exposure to ambient environment can rapidly degrade the performance of the product.

Identifiers

code_infoCatalog 340334, Lot ;6083708 (embossing ID of (160216T3)